Dylix Corporation DPL Series
Process Differential Pressure Transmitters

Characteristics:
- 316 SST diaphragms and flanges
- External zero and span adjustments
- ±0.2% STR Static Accuracy
- Hasteloy C-276/ Monel diaphragms
- Hart Protocol
- Local display


Designed for many markets/applications.
Dylix’s DPL Series DP is a high static—low differential
pressure transmitter based on proven capacitance sensing
technology. Key feature of the sensing element is its
isolation for induced stress which improves performance and
long term stability over elevated line pressures.
Manufactured for extended life cycle.
Reliability and stability are manufactured into every device
using advanced manufacturing techniques. Extended burnin
and thermal cycling are just a few extra steps Senex utilizes
to ensure performance over time.
Entirely welded 316L SST pressure cavities.
The DPL comes standard with 316L SST sensing diaphragms
and process flanges to handle most aggressive fluids.
Hasteloy C276 is optional.
General Specifications :

Electrical | |
Power Supply |
12 to 45 Vdc with no load
|
Output |
4-20 mAdc linear
|
Zero and Span Adjustment: |
Continuous
|
Warm up: |
2 seconds max (with minimum damping)
|
Damping: |
Variable from 0.2 to 1.67 seconds
|
Power Supply Effect: |
Less than +/-0.005% of span/volt input change
|
Load Effect: |
No effect except power across transmitter
|
EMI/RFI: |
ILess than +/-0.1% FSO;20 to 1000 MHz; @ 30 V/m
|
Short Circuit : |
Protected Up to 40 Vdc
|
Performance | |
Static Accuracy: |
±0.5% FSO* (BFSL, RSS)(combined effects of non-linearity, hysteresis &repeatability)
|
Static Pressure Effect: |
On zero =±0.25% FSO per 1,000 psiOn span =±0.5% of reading per 1,000 psi
|
Repeatability : |
±0.05% FSO
|
Temperature Effects : |
±1.5% FSO over comp range(combined effects of Zero & FSO with reference at 70 oF)
|
Long Term Stabily: |
±0.25% FSO per year
|
Pressure Ranges | |
Standard Ranges (STR): |
0-30”, 150”, 750” H2O
|
Range ability: |
6:1 max
|
Maximum Static Pressure: |
2000 psi
|
Overpressure Protection: |
2000 psi without changing calibration
|
Burst Pressure: |
8,700 psi
|
Performance | |
Static Accuracy: |
+/-0.2% of calibrated span
|
Base Pressure EffectOn ZeroOn span |
+-0.25% STR of standard range per 2000 psi+/-0.5% of reading for 0-30” H2O+/-0.25% of rdg per 1000 psi; 0-150” & 750”H2O
|
Temperature Effect: |
+/-2.5% STR Total Error Band for 0-30”H2O+/-1.5% STR Total Error Band; 0-150” & 750”H2O
|
Vibration Effect: | 0.05% STR per g to 200 Hz in any axis |
Mounting Effect: | Zero shifts up to 1” H2O depending on orientation |
Stability: |
+/-0.2% STR per 6 months
|
Mechanical | |
Wetted Parts :‘0’-Rings |
Diaphragms:
316L SSTOptions: Hasteloy C-276, Monel,Process Flanges: 316L SSTViton |
Non-Wetted Parts: |
Bolts:
Plated carbon steelElectrical Housing: Low-copper aluminumPaint: PolyurethaneCover ‘O’-Ring: Buna-NFill Fluid: Silicon Oil |
Prces Connection : |
¼”-NPT Female*
|
Port Centers : |
2 1/8”
|
Pressure Port: |
¼”-NPT Female*
|
Electrical Connector : |
½-14 NPT Female w/ screw
terminals |
Weight: | Approximately 6.5 |
Environmental | |
Temperature Limits Storage |
-60 to + 250 °F
|
Electronics: |
-40 to 200 °F
|
Sensing: |
-40 to 250 °F
|
Volumetric Displacement: | Less than 0.01 in3 |


![]() ![]() |
|||||||||||||||||||||
|