Yokogawa Model CA11 Voltage/Current Calibrators |
||
![]() |
Product Information:
Product Description:
Features:
|
![]() |
Order YOKOGAWA TMI Products on-line |

Product Specifications:
- Source and Measuring of Voltage and Current:
General and measures voltages up to 30V DC and current up to 24 mA DC - 4-20 mA and 1-5 VDC Step-up/down Function:
- The output level can be changed between 1↔2↔3↔4↔5V signals for the 1 to 5V DC output, for efficient calibration work.
- 20 mA SINK Function:
Absorb the voltage supplied from an external power supply to its H terminal and simulates a two-wire transmitter, idea for loop checks. - Sweep Function:
Increase and decrease the output level to the preset level at a constant rate for the selected sweep time (16 or 32 seconds). The sweep function and sweep time are set by the internal dip switch.
Source Function:

Signal Range | Accuracy | Resolution | |
30V |
0 to 30.00 V DC
|
0.05% + 20 mV | 10 mV |
10V | 0 to 11.000 V DC | 0.05% + 2 mV | 1 mV |
1-5V | 1/2/3/4/5 V DC | 1-V step | |
1V | 0 to 1.1000 V DC | 0.05% + 0.2 mV *1 | 0.1 mV |
100mV | 0 to 110.00 mV DC | 0.05% + 50 μV *1 | 0.01 mV |
20mA | 0 to 24.00 mA DC | 0.05% + 4 μA | 0.01 mA |
4-20mA |
4/8/12/16/20 mA DC
|
4-mA step | |
20mA SINK |
0.1 to 24.00 mA DC
|
0.1% + 4 μA | 0.01 mA |
Temperature effect: 1/10 of accuracy/°C; however,
for 100-mV range, 0.005% + 10 μV/°C Measurement Function:
Signal Range | Accuracy | Resolution | |
30V |
0 to ±30.00 V DC
|
0.05% + 3°C 0.07% + 3°C | 10 mV |
10V | 0 to ±11.000 V DC | 0.05% + 2°C 0.07% + 2°C | 1 mV |
1V | 0 to ±1.1000 V DC | 0.05% + 30 μV 0.05% + 30 μV |
0.1 mV |
100mV | 0 to ±110.00 mV DC | 0.05% + 0.6°C *6 0.05% + 0.6°C *7 |
0.01 mV |
20mA | 0 to ±24.00 mA DC | 0.05% + 0.2 W *6 0.05% + 0.2 W *7 |
0.01 mA |