- Literature
- Specifications
- Features/Options
- Modules
- Accessories
Model: Yokogawa SL1400 ScopeCorder |
Models |
SL1400 |
Input type: |
Plug-in module (A/D converters built in to each unit) |
Number of module slots |
8 |
Sample rate |
10 MS/s (Maximum sample rate differs depending on the type of module.) |
Max. Record length |
2.5 MW/CH, 50 MW total |
Channels |
16 channels + 16-bit logic (8 bits x 2)
* Dependent on modules and no. of channels |
Time axis accuracy |
±0.005% |
Time axis setting |
100 μsec to 5 sec (in steps of 1, 2, or 5)
10 sec to 60 min (in steps of 1-2-3-5-6)
100 min/120 min/300 min
10 h/20 h/30 h/50 h/100 h/120 h
10 days/20 days/30 days |
External clock input |
Connector type: RCA jack, Input level: TTL level (0 to 5 V), Frequency range: 1 MHz or less |
Display—- |
Display: |
10.4-inch TFT color LCD monitor, res. 800 x 600 |
Display res. of waveform display: |
650 × 512 (normal waveform display) or 750 × 512 (wide waveform display) selectable |
Display format: |
Zoom – MAIN, MAIN&Z1, MAIN&Z2, MAIN&Z1&Z2, Z1only, Z2only, and Z1&Z2
XY -TY, XY, and TY&XY
1, 2, 3, 4, 8, or 16 analog waveform windows |
Functions |
Acquisition
mode: |
Normal |
Maximum sample rate of 10 MS/s |
Envelope |
Maximum sample rate regardless of record time, holds peak value |
Averaging |
Average count 2 to 65536 (2n steps) |
Box (av.) |
Increase A/D resolution up to 4 bits (max 16 bits) |
Roll |
Roll display for recording time of 1 sec or more |
Zoom: |
Expand the displayed waveform along the time axis (up two locations using separate zoom rates) |
Display format: |
2, 3, 4, 8, or 16 analog waveform windows |
X-Y display: |
Select the X axis and Y axis from CH1 toCH16, MATH1 to MATH8 (up to 4) |
Snapshot: |
Retains the current displayed waveform on the screen. Snapshot waveforms can be saved and loaded. |
Accumulation: |
Accumulates waveforms on the display (persistence mode) |
Analysis |
Auto scroll: |
Automatically scrolls the zoom position. |
History search function: |
Search for and display waveforms from the history memory that satisfy specified conditions. Zone search/Parameter search |
Cursor measurement: |
Horizontal, Vertical, H&V, Degree (only for T-Y waveform display), and Marker |
Vertical/Horizontal Axis Settings |
Roll mode: |
The roll mode is enabled when the trigger mode is set to auto, auto-level, single, or log, and the display span along the time axis is greater than or equal to 1 s. |
Linear scaling: |
Set AX+B mode or P1-P2 mode independently for CH1 to CH16. |
ALL CH menu |
Set all channels while displaying waveforms. Operation using the USB keyboard and USB mouse is possible. |
Data Storage |
History memory |
Automatically holds up to 2000 pages of waveforms (depending on the memory length |
Storage |
Saves waveform data, setup data, snapshot waveform data, the results of the automated measurement of waveform parameters, and screen image data |
Real Time Recording (Recorder Mode) |
Real time recording on the built-in printer: |
T-Y waveform recording/numeric value recording: Output to the chart in real time
X-Y waveform recording: Starts data acquisition with START and generates X-Y
waveforms in real time
Outputs X-Y waveforms to the chart with STOP. |
Length of data saved to memory while real time recording: |
T-Y waveform recording: Fixed to 2.5 MW Automatically saves up to 1000 divisions of data (depending on the chart speed).
X-Y waveform recording: Fixed to 1 MW |
Recording Format |
T-Y waveform recording: Select from 1, 2, 3, 4, 8, or 16 division recording (flexible zone is selectable for 1 division recording)
Numeric value recording: Print direction selectable from standard and 180° rotation.
X-Y waveform recording: Records up to 4 waveforms simultaneously. Assignment of X and Y channels on the 4 waveforms is arbitrary. |
Built-in Printer (/B5 Option) |
Print type: |
Thermal line/dot matrix |
Paper width: |
210 mm |
Effective print width: |
204 mm (1632 dots) |
Feeding direction res.: |
8dot/mm |
PC Card Interface |
Slots |
2 (front panel (1), rear panel (1)) |
Supported card |
GPIB card (National Instruments NI PCMCIA-GPIB card),
Flash ATA memory card (PC card TYPE II), CF card + adapter
card, and various hard disk type PC cards |
Storage |
Built-in storage: |
PC card interface (optional) |
External storage interface |
SCSI/USB storage device |
Built-in HDD |
40 GB |
Environment |
Power supply: |
100 to 120VAC/220 to 240VAC (automatic switching) — 50/60Hz |
Power consumption: |
Max 200 VA |
Operating Temperature: |
5C° to 40C° |
Dimensions: |
355 mm (W) x 250 mm (H) x 225 mm (D) |
Weight: |
Approx. 7.8 kg (only the SL1400 with all options (/C8/C10/P4 options)
Approx. 11.8 kg (SL1400 + 701250 x 8) |
Accurately Capturing Complex Signals or Long Waveforms, Using the SL1400 large Memory Capacity
The ScopeCorder’s standard memory capacity is 50 MW (2.5 MW per channel)
- GigaWord Recording
—You can record data for 10 days (1 day/div) on the main screen, while displaying 1-second recordings (100 ms/div) in real time on the zoom screen.
The large memory capacity lets you capture all of your data while still maintaining a sample rate fast enough to see any abnormal phenomena.
- Efficient Memory Use
— Sufficient memory length is available even when 16 channels are used, so you can conduct extended observations on multiple channels (2.5 MW per channel with standard memory)
Automatically Measure Waveform Parameters
Waveform parameters such as voltage, frequency, and RMS are measured automatically. In addition to general parameter measurement function, the ScopeCorder comes standard with
functions such as the following:
- Easily Find and Display Waveform Frequency, Rise Time, and Other Parameter
This function calculates statistical information about the waveform.
Maximum value, minimum value, average value, and standard deviations are calculated automatically for each waveform parameter. In addition, you can instantaneously search for the
cycle containing the maximum value and display it on the zoom screen. This cycle statistical calculation greatly improves your insight enabling you to analyze transient phenomena
captured using the long recording memory
Snapshot Function
- Enables On-Screen Waveform Comparisons
Using the snapshot function, you can keep the currently
displayed waveform with the touch of a button. Snapshots
are useful for comparing a reference waveform with an input
waveform. In addition, snapshots can be saved to and loaded
from the storage media.
Memory Backup Function
- — This function backs up about 150 hours of data saved to the acquisition memory immediately backup helps you avoid losing important data even if the power
supply is unstable and gets cut off.
(Backup time varies according to the usage environment. Four AA batteries are required for memory backup.)
PDF output for printing A4-sized reports
- — When performing on-site measurements, you can print out the data and hand-write memos on the paper. Since the SL1400 simultaneously stores data to internal memory while printing, you can keep electronic and hard copy records with just a single action. (Remember that with thermal-sensible paper, it is vital to make photocopies for long term preservation.) The SL1400 allows you to export results to files in PDF format, making it easy to save data for long periods of time, transfer the data to distant locations, or load them onto a PC. It is also easy to create reports since waveform data can be converted to an A4-size layout.
Real-Time Hard Drive Recording (with the /C8 Option)
With the optional internal hard drive, you can record measurements
to the hard drive in real time.
This makes it easy to manage and
analyze data using a PC.
- Maximum data capacity:
1GW
- Maximum sampling rate:
100 kS/s
* using 1 Channel only
Ethernet Interface (/C10 Option)
Connecting to a PC as a Web Server and FTP Server
- — The SL1400 has a variety of server functions that let you perform remote controls or download waveform data and screen images onto a PC. You can also access the SL1400 using
Internet Explorer. Same as with RS232 and GB-IB, you can write your own custom programs in Visual C++ 6.0 or Visual Basic 6.0 and control the SL1400 through an Ethernet interface
Web Server Functions
— Connect the SL1400 to your PC through the Ethernet connection. This allows for easy remote operation using Internet Explorer.
- FTP
You can easily copy and paste files to and from a PC and the instrument’s flash memory or other storage media.
- Measurement Trend
Using Internet Explorer, you can periodically or manually download screen images to a PC for remote waveform monitoring. You can also download waveform
data, start or stop a measurement, or setup a split display all from a PC.
- Data Capture
This function downloads values of waveform parameters periodically, launches MS Excel and graphs the parameters on a spreadsheet values. This enables you to check the parameter
trends at a glance.
Probe Power, 4-output (/P4 Option)
Connect the SL1400 to your PC through the Ethernet connection. This allows for easy remote operation using Internet Explorer.
- Maximum number of probes powered : 4
- Compatible probes :
Current probes 701933 (30 A) & 701930 (150 A)
- Up to 2 current probes can be used at one time
|
Volts |
- 701250 – High-Speed 10 MS/s, 12-Bit Isolation Module
- 701251 – High-Speed 1 MS/s, 16-Bit Isolation Module
- 701255 – High-Speed 10 MS/s, 12-Bit Non-Isolation Module
- 701260 – High-Voltage 100 kS/s, 16-Bit Isolation Module
|
|
Temp. Volts |
- 701261 – Universal (Voltage/ Temp.) Module
- 701262 – Universal (Voltage/ Temp.) Module with AAF
- 701265 – Temperature, High Precision Voltage Isolation Module
|
|
Strain |
- 701270 – Strain Module (NDIS)
- 701271 – Strain Module (DSUB, Shunt-Cal)
|
|
Volts |
- 701275 – Acceleration/Voltage Module (with AAF)
|
|
Hz |
- 701280 – Frequency Module
|
|
|
Model |
Sample Rate |
Res. |
|
Ch |
Max. V |
Acc |
Volts |
•701250 |
10 MS/s |
12-Bit |
3 MHz |
2 |
600 V*2
250 V*3 |
±0.5% |
high noise immunity |
•701251 |
1 MS/s |
16-Bit |
300 kHz |
2 |
>600 V*2
140 V*3 |
±0.25% |
High sensitivity range (1mV/div), low noise(±100 μV typ.), and high noise immunity |
•701255 |
10 MS/s |
12-Bit |
3 MHz |
2 |
600 V*4
250 V*3 |
±0.5% |
non-isolation version of model 701250 |
•701260 |
100 kS/s |
16-Bit |
40 kHz |
2 |
1000 V*2
850 V*3 |
±0.25% |
with RMS, and high noise immunity |
Temp.
Volts |
•701261 |
100 kS/s (V),
500 S/s (Temp) |
16-Bit (V),
0.1: (Temp) |
40 kHz (V),
100 Hz (Temp) |
2 |
42 V |
±0.25% (V) |
thermocouple (K, E, J, T, L, U, N, R, S, B, W, iron-doped gold/ ) |
•701262 |
100 kS/s (V),
500 S/s (Temp) |
16-Bit (V),
0.1: (Temp) |
40 kHz (V),
100 Hz (Temp) |
2 |
42 V |
±0.25% (V) |
thermocouple (K, E, J, T, L, U, N, R, S, B, W, iron-doped gold/chromel), with AAF |
•701265 |
500 S/s (V),
500 S/s (Temp) |
16-Bit (V),
0.1: (Temp) |
100 Hz |
2 |
42 V |
±0.08 (V) |
thermocouple (K, E, J, T, L, U, N, R, S, B, W, iron-doped gold/chromel), high sensitivity range
(0.1mV/div), and low noise (±4 μV typ.) |
Strain |
• 701270 |
100 kS/s |
16-Bit |
20 kHz |
2 |
10 V |
±0.5% (Strain) |
Supports strain NDIS, 2, 5, 10 V built-in bridge power supply |
• 701271 |
100 kS/s |
16-Bit |
20 kHz |
2 |
10 V |
±0.5% (Strain) |
Supports strain DSUB, 2, 5, 10 V built-in bridge power supply, and shunt CAL |
Analog
Voltage
Accl. |
• 701275 |
100 kS/s |
16-Bit |
40 kHz |
2 |
42 V |
±0.25% (V)
±0.5% (Acc.) |
built-in anti-aliasing filter, Supports built-in amp type acceleration sensors
(4 mA/22 V) |
Hz |
• 701280 |
25 kS/s |
16-Bit |
res. 50 ns |
2 |
420 V*2
42 V*3 |
±0.1% (Hz) |
Measurement frequency of 0.01 Hz to 200 kHz, Measured parameters (frequency, rpm, period,
duty, power supply frequency, distance, speed) |
*1: Probes are not included with any modules.
*2: In combination with 10:1 probe model 700929
*3: Direct input
*4: In combination with 10:1 probe model 701940
*5: Some of the models 701250/701255 shipped on or before July, 2007 may require factory rework. |
STANDARD ACCESSORIES |
|
- Passive Probes
- Differential Probes
- Current Probes
- 10:1 Probes
- Bridge Head Probes
- Logic Probes
- Software
- Miscellaneous- Leads
|
OPTIONAL ACCESSORIES (Brochure Download) |
Passive Probes |
 |
701940 — 10 MΩ (10:1), 10 MHz, 1.5 m — Passive probe Passive Probe |
Differential Probes |
 |
700924 — DC up to 100 MHz (-3 dB) bandwidth/4MΩ/max. — Differential Probe Differential probe capable of wide-band, high-voltage floating measurements |
Current Probes |
 |
701930 — DC to 10 MHz bandwidth, 150 Arms — Current probe Bandwidth: DC to 10 MHz (-3dB) |
 |
701931 —DC to 2 MHz bandwidth, 500 Arms — Current probe Bandwidth: DC to 2 MHz (-3dB) |
 |
701933 — DC to 50 MHz BW, 30 Arms — Current probe Bandwidth: DC to 2/10/50/100 MHz; current measurement: Max 30/150/500 A
1 MΩ BNC inputs without the need for an external amplifier |
10:1 Probes |
 |
700929 — DC to 100MHz (-3dB /1,000V (DC + AC peak, probe itself)— Isolated BNC probe 10:1 Probe |
Bridge Head Probes |
 |
701955 —(NDIS, 120 Ω)—Bridge Head Probe Bridge resistance: 120Ω |
 |
701956 —(NDIS, 350Ω)—Bridge Head Probe Bridge resistance: 350Ω |
 |
701957 —(Dsub, 120 Ω)—Bridge Head Pro bee D-sub cable (5 m) included. Supports Shunt-Cal. |
 |
701958 —(Dsub, 120 Ω)—Bridge Head Probe D-sub cable (5 m) included. Supports Shunt-Cal. |
Logic Probes |
|
702911 — Logic Probe
8-Bit, 1 m, non-Isolated, TTL level/Contact Input, Includes one each of the B9879PX and B9879KX connection leads. |
|
702912 — Logic Probe
8-Bit, 3 m, non-Isolated, TTL level/Contact Input, Includes one each of the B9879PX and B9879KX connection leads. |
|
700986 — 100kΩ, ±42 V (DC +AC peak) 1.5m, Non-isolated— Logic Probe
8-bit, non-isolated, response speed: 1µs, two measurement leads (B9879PX and B9879KX) included |
|
700987 — 100kΩ, ±42 V (DC +AC peak) 1.5m, isolated— Logic Probe
8-Bit, each channel isolated |
Software |
 |
701992 —Xviewer— For DL series Software |
 |
701992-62E —Xviewer— For DL advanced series Software |
Miscellaneous |
 |
701934 PROBE POWER SUPPLY A power supply for current probes, FET probes, and differential probes. Supplies power for up to four probes, including large current probes. |
|
701901 – 1:1 Safety BNC Adapter Lead(in combination with following)
1000 Vrms-CAT II (BNC-BNC)
701959— Safety Mini-Clip (Hook type) 1000 Vrms-CAT II, red and black set
701954— Large Alligator-Clip (Dolphin type) 1000 Vrms-CAT II, red and black set
758929— Alligator Clip Adaptor Set (Rated Voltage 1000 V) 1000 Vrms-CAT II, red and black set
758922— Alligator Clip Adaptor Set (Rated Voltage 300 V)300 Vrms-CAT II, red and black set
758321— (Fork Terminal Adapter Set) 1000 Vrms-CAT II, red and black set
|
|
701902 —Safety BNC-BNC Cable (1 m)
1000 Vrms-CAT II (BNC-BNC) |
|
701903 —Safety BNC-BNC Cable (2 m)
1000 Vrms-CAT II (BNC-BNC) |
|
758924 — Safety BNC-banana Adapter
500 Vrms-CAT II |
|
758917 — Measurement Lead Set
Measurement leads (2 per set) Alligator-Clip is required separately. |
|
366926 — 1:1 BNC-Alligator Cable
Non-isolated 42 V or less, 1m |
|
366961 — 1:1 Banana-Alligator Cable
Non-isolated 42 V or less, 1.2m |
|
B9988AE —Printer Roll Paper |
|
701967 — Soft Carrying Case
for SL1400 |
|
|